PVCHECK
EN - 45
6.4.4.1. Anomalous situations
1. In any operating mode, in case the instrument detects a
voltage higher than 1000V at terminals P-N, P-E and N-
E, it does not carry out the test, gives out a long sound
and displays the message “Vin > 1000”.
15/05/12 15:34:26
Ins. Test : 1000 V
Ri min : 1.0 M
Mode : String
Vtest = - - - V
Rp = - - -
M
Vin > 1000
Selection
M
2. In STRING mode, in case the instrument detects an Isc
current higher than 10A, it does not carry out the test,
gives out a long sound and displays the message “Isc
current too high”.
15/05/12 15:34:26
Ins. Test : 1000 V
Ri min : 1.0 M
Mode : Stringa
Vtest = - - - V
Rp = - - -
M
Isc current too high
Selection
M
3. In STRING mode, in case the instrument detects a
current < 0.2A between terminals P and N, it does not
carry out the test, gives out a long sound and displays
the message “Current < Lim”.
15/05/12 15:34:26
Ins. Test : 1000 V
Ri min : 1.0 M
Mode : String
Vtest = - - - V
Rp = - - -
M
Current < Lim
Selection
M
4. In STRING mode, in case the instrument detects a
voltage < 15V between terminals P and N, it does not
carry out the test and displays the message “Low
voltage”.
15/05/12 15:34:26
Ins. Test : 1000 V
Ri min : 1.0 M
Mode : String
Vtest = - - - V
Rp = - - -
M
Low voltage
Selection
M