5) Carry out the measurement by pressing the START/STOP button. Test result will be displayed
after elapsing measurement time in green colour equipped with green
symbol and with beep-
beep sound if it is lower than or equal to limit value (see the explanation of limit value above). If the
result is higher than limit value, then it will be displayed in red colour equipped with red symbol
and with longer beep sound. See the display outlook with test result on the figure 27.
6) Save the test result by pressing the SAVE hard key twice, for further instructions see the
“MEMORIZING EXAMPLE” section.
7) Press the EXIT hard key to clear displayed result, actual input voltage UL/PE will start to be
displayed again.
CAUTION
In case voltage UL/PE within required range 100 265 V is present between L
and PE test terminals (also displayed) but there is no READY message
displayed, check if energizing socket is correctly earthed!
The following specific information can be shown on the display during the measurement:
Information displayed Description
VOLTAGE OUT OF RANGE!
Input voltage UL/PE out of required range 100 265 V
after pressing the START key.
CONTACT VOLTAGE > 50 V
or
CONTACT VOLTAGE > 25 V
Contact voltage higher than selected limit value, probably
due to too high loop resistance.
MEASUREMENT FAILED!
Measurement current was interrupted either due to
disconnection of test leads or due to increase of loop
resistance.
FUSE F3!
Fuse F3 is blown.
HOT!
Internal circuitry is overheated.
Wait to cool it down!